Byond Infusion Pump BYS–820 is a bubble detector and pressure sensor that can precisely detect air bubbles. It features KVO and Bolus functions, a unique voice alarm system and an accurate injection rate to make the injection process safer and more reliable.
Features:
- HD LCD Display, high capacity words, friendly user interface, dynamically display working status;
- Audible and visual alarm for occlusion, empty, low battery, end of infusion, door open, wrong setting, etc., which gain patents;
- Compatible with any brands of infusion sets after correct calibration;
- Preset solution volume to greatly reduce the workload of nurses;
- Work mode: ml/h and drop/min can switch freely;
- Three levels of occlusion: high, middle and low;
- Purge function;
- KVO (keep-vein-open) automatically opens as infusion is completed, KVO rate is 1-5ml/h(1ml/h step);
- Power Source: AC100—240V, 50/60Hz; Internal Battery;
- Automatically record the settings of last infusion;
- OEM is available.
Specifications:
Rate | 1ml/h ~ 1,200ml/h |
Flow Rate Accuracy | Within +/- 5% (after correct calibration) |
Mechanical precision | Within +/- 2% |
Purge Rate | 100ml/h ~ 1,000ml/h (100 ml/h step) |
Infusion Volume | 1ml ~ 9999ml |
Total Infusion Volume | 0.1ml ~ 9999.9ml |
KVO rate | 1ml/h ~ 5 ml/h (1ml/h step) |
Occlusion | High: 800mmHg +/- 200mmHg (106.7kPa +/- 26.7kPa) Medium: 500mmHg +/- 100mmHg (66.7kPa +/-13.3kPa) Low: 300mmHg +/-100mmHg (40.7kPa +/-13.3kPa ) |
Unique audible and visible alarm | Human voice alarm for end of injection, occlusion, door open, bubbles in the tube, wrong setting, low battery, AC power pulled out etc. |
Power Source | AC 100V ~ 240V, 50/60Hz; Internal rechargeable Li battery, capacity1,600mAh, 4 hours internal battery backup |
Bubble detector | Ultrasonic wave detector; detection sensitivity 25mu/L |
Fuse | F1AL/250V, 2 pc inside |
Power Consumption | 18VA |
Operating Condition | Ambient temperature: +5℃ ~ +40℃; Relative humidity: 20 ~ 90% Atmospheric pressure: 86.0kpa ~ 106.0kpa |
Transport & Storage Condition | Ambient temperature:-30℃ ~ +55℃ Relative humidity: ≤95% |
Equipment Classification | Class II, internal power supply, Type CF |
IP Classification | IPX4 |
Dimension | 140mm (L);157mm(W);220mm(H) |
Weight | 1.8kg (net weight) |
In the box:
- Infusion Pump
- Operating Manual
- Power Cord
Additional information
Warranty | 1 year |
---|
Reviews