| PARAMETER |
Paediatric
|
Adult
|
Adjustments
|
|---|---|---|---|
| Pressure slope |
1, 2, 3, 4
|
1, 2, 3, 4
|
– |
| Inspiration pressure (cm H₂O) |
3-40 in NIV
3-60 in Invasive
|
5-60
|
1 cm H₂O
|
| Pressure support (cm H₂O) |
3-40 in NIV
3-60 in Invasive
|
5-60
|
1 cm H₂O
|
| PEEP (cm H₂O) |
0–20
|
0–25
|
1 cm H₂O
|
| Tidal volume (mL) |
50-500
|
300-2500
|
10 mL (50-1000)
100 mL (1000-2500) |
| Adjustable plateau time (seconds) | 0–1.5 | 0–2.0 | 0.1 sec |
| Maximum flow setting (volume) (L/min) |
5-40
|
60 (in volume modes), 220
(in pressure modes) |
1 L/min
|
| Flow shape |
(1-5) Square/Decreasing
|
(1-5) Square/Decreasing
|
–
|
| I/E ratio (1/x) |
1/0.4-1/9.9
|
1/0.40-1/9.9
|
1/0.1
|
| Respiratory rate |
2-80
|
2–50
|
1 bpm
|
| FiO2 (%) | 21–100 | 21–100 | 5% (30–100) |
| Maximum inspiration time (seconds) |
1.0–3.0
|
1.0–3.0
|
0.1s
|
| Apnea time (seconds) |
5-20
|
10-60
|
1s
|
| TRIGGER | |||
| Inspiration pressure trigger (cm H₂O)
Invasive ventilation only) |
0.2–5.9 / NO in (A)CV and P(A)CV modes
0.2–6.0 in PS.SV, PS.Vt, SIMV and PSIMV modes |
0.2–5.9 / NO in (A)CV and P(A)CV modes
0.2–6.0 in PS.SV, PS.Vt, SIMV and PSIMV modes |
0.1 cm H₂O
|
| Inspiration flow trigger (L/min)
(Invasive ventilation only) |
0.2–9.9 / NO in (A)CV and P(A)CV modes
0.2–10.0 in PS.SV, PS.Vt, SIMV and PSIMV modes |
0.2–9.9 / NO in (A)CV and P(A)CV modes
0.2–10.0 in PS.SV, PS.Vt, SIMV and PSIMV modes |
0.1 L/min
|
| Inspiration trigger
(Noninvasive ventilation only) |
AUTO / 1–5 / NO in (A)CV and P(A)CV modes
AUTO / 1–5 in PS.SV, PS.Vt, SIMV and PSIMV modes (NO for E250) |
AUTO / 1–5 / NO in (A)CV and P(A)CV modes
AUTO / 1–5 in PS.SV, PS.Vt, SIMV and PSIMV modes (NO for E250) |
–
|
| Expiratory trigger
(% peak flow) |
10 / 90 AUTO in PS.SV, PS.Vt, SIMV and PSIMV modes
|
10 / 90 AUTO in PS.SV, PS.Vt, SIMV and PSIMV modes
|
1 %
|
| RECRUITMENT / SIGH | |||
| Period (min) |
NO / 1-60
|
NO / 1-60
|
1 min
|
| Duration (seconds) |
0.2-40
|
0.3-40
|
0.1 sec (0.2–3) 1 sec
(3–40) |
| Pressure (cm H2O) |
5–60
|
5-60
|
1 cm H₂O
|
| NEBULISATION | |||
| Period (hr) |
1–24
|
1–24
|
1 hr
|
| Duration (min) | 1–30 | 1–30 | 1 min |
| Flow (L/min) | 5–20 | 5–20 | 1 L/min |
| Size |
26 cm x 24 cm x 13 cm (10.2” x 9.4” x 5.1”)
|
| Weight |
9.9 lb (4.5 kg) with mains supply pack
|
| Internal battery |
3 hrs*
|
| External battery |
3 hrs*
|
| Modes: |
PAC, PS, PS.VT, ACV, CPAP, PSIMV, VSIMV
|
FEATURES :
- High mobility and uninterrupted ventilation
- Transportability
- Safe and secure
- Ready to go
- Advanced performance
- Compact and easy to use

Philips Trilogy EVO (OBM)
Micomme P1 BIPAP with Humidifier
Resmed Lumis 100 VPAP S BIPAP
OxyMed SleepEasy Auto CPAP
Resmed Airsense 11 Autoset CPAP
ResPro ST25A Auto Bipap machine with AVAPS and Humidifier
Breas Vivo 65
Noccarc V310 With Integrated HFNC 

Reviews
There are no reviews yet.